NCAR Recall
Back
Reference Number:
mdprc 036 02 13 000
Date submitted:
2/6/2013
Manufacturer:
Radiometer Medical
Device Type:
Analyser, ABL90 FLEX sensor cassette
Description:
Analyser, blood gas, general purpose
Medical Device Identifier:
Lot/batch no.: R0124N088 to R0124N129
Reason of Field Safety Corrective Action:
The expiration date for the ABL90 FLEX SC is 4 months from the production date, including the 30 day in use lifetime . SC90 946-008R0124 (42 cassettes #88-129) were produced on 2012-10-23, while the labels were erroneously printed with the expiration date 2013-10-23. The chip in the SC is encoded with the date from the barcode label, and is therefore also incorrect
Remedy Action:
Radiometer has requested the affected customers to return any sensors in stock or in use
Athorized Representative/Importer/Distributor:
Salehiya Trading Est, Jeddah, 01-2628939
Report Source:
NCAR
Source Ref. Number:
DK-2013-01-31- 023
SFDA Comments:
SFDA urges all hospitals that have devices subjected to recall, to contact the company.