Reference Number:
mdprc 081 03 12 000
Date submitted:
3/20/2012
Manufacturer:
Prism Medical Ltd
Device Type:
Freeway Mobile Hoists, SPREADER BAR
Description:
Hoists, Mobile, SPREADER
Medical Device Identifier:
Freeway Mobile Hoists FME150, FME175, FME200, FMH150, FMH175 & FMH200 Serial no.: FGE, FSE, FME, FGH, FSH and FHM between serial numbers 1500 & 6500
Reason of Field Safety Corrective Action:
A hoist spreader (carry) bar became detached during a patient transfer. The investigation identified that a bearing washer had been omitted during the assembly process of a carry bar
Remedy Action:
The manufacturer has issued a Field Safety Notice to inform users that an inspection is required on the mounting pin of the spreader bar.
Athorized Representative/Importer/Distributor:
N/A
Report Source:
NCAR
Source Ref. Number:
UK-2012-03-14-043
SFDA Comments:
SFDA urges all hospitals that have devices subjected to recall, to contact the company.