Reference Number:
mdprc 051 01 16 000
Date submitted:
1/13/2016
Manufacturer:
Bio Rad.
Device Type:
D-10 Rack Loader
Description:
allows loading up to 50 samples for walk-away capability
Medical Device Identifier:
Catalog No. 220-0600
Reason of Field Safety Corrective Action:
It has come to our attention that there is a potential to assign a patient result to an incorrect sample ID when running in the D-10 Rack Loader configuration.
Remedy Action:
While reviewing results, please verify that the number of results equals the total number of samples run. If there is a discrepancy, identify the rack with the missing result, rerun all samples from that rack, and check the laboratory information system (LIS) results.
Athorized Representative/Importer/Distributor:
Al-Jeel Medical & Trading Co. LTD
Report Source:
NCMDR
Source Ref. Number:
05687C1862163
SFDA Comments:
SFDA urges all hospitals that have devices subjected to recall, to contact the company.