NCMDR Recall
Back
Reference Number:
mdprc 063 06 14 000
Date submitted:
6/18/2014
Manufacturer:
Johann Stiegelmeyer & Co GmbH
Device Type:
Transport Stretcher
Description:
Beds
Medical Device Identifier:
Model: Mobillo Type: 201412 Number: 208730; 222328; 255681; 270907
Reason of Field Safety Corrective Action:
Possibility of instability of the mattress support platform because of defective gas springs of the tilting mechanism in the undercarriage due to heavy duty use/ overload.
Remedy Action:
The gas spring will be replaced by ratchet tubes, which will increase resistance to the operating forces to optimal design parameters.
Until the corrective action is made, users are advised not to overload the stretcher and to avoid unbalanced loads on the ends (head-/ foot end of the mattress support platform).
Athorized Representative/Importer/Distributor:
Al Hammad Medical Services, (011) 4659811
Report Source:
NCMDR
Source Ref. Number:
792EFAC5502B6
SFDA Comments:
SFDA urges all hospitals that have devices subjected to recall, to contact the company.