Reference Number:
mdprc 002 07 11 000
Date submitted:
7/3/2011
Manufacturer:
Intuitive Surgical Inc
Device Type:
da Vinci Intuitive Surgical System
Description:
Telemanipulation
Systems, Surgical,
Minimally Invasive
Medical Device Identifier:
da Vinci Si Surgical System (IS3000). Serial no: SH0565
Reason of Field Safety Corrective Action:
This field correction is being initiated due to failure of a retention component in the Master Tool Manipulator (MTM – commonly referred to as the “Master”) spring counterbalance subsystem.
Remedy Action:
The surgeon should always assure that he/she does not release the grip of the masters while his/her head is in the stereo viewer. Intuitive Surgical has designed reinforcing caps for the retention component to correct the problem with the MTM subsystem.
Athorized Representative/Importer/Distributor:
Gulf Medical Co. Jeddah (02) 6512828
Report Source:
NCMDR
Source Ref. Number:
A7D8F7B5B93E4
SFDA Comments:
SFDA urges all hospitals that have devices subjected to recall, to contact the company