NCMDR Recall
Back
Reference Number:
mdprc 092 01 13 000
Date submitted:
1/23/2013
Manufacturer:
GE Healthcare
Device Type:
Infant radiant warmers, Lullaby Warmers
Description:
Infant radiant warmers
Medical Device Identifier:
All Lullaby Warmers (Model # 2041599-001)
Reason of Field Safety Corrective Action:
If after fully assembling the Lullaby Warmer, it has been transported under rough conditions such as outdoors, in a truck, or over difficult/abrupt thresholds, there is a potential for the joint between the heater head and the support structure to break. If the breakage goes undetected, the head may fall on a patient resulting in a severe adverse patient outcome.
Remedy Action:
To minimize any safety risk associated with the Lullaby Warmer, stop using the unit immediately until the issue is corrected. A GE Healthcare representative will contact you to make arrangements to install the correction when it becomes available.
Athorized Representative/Importer/Distributor:
GE Healthcare, Riyadh, (01) 4600530
Report Source:
NCMDR
Source Ref. Number:
60E603E06428D
SFDA Comments:
SFDA urges all hospitals that have devices subjected to recall, to contact the company.