NCMDR Recall
Reference Number: mdprc 160 09 12 000
Date submitted: 9/25/2012
Manufacturer:Radiometer Medical
Device Type:IVD, analyzer ABL90 FLEX
Description:IVD, analyze
Medical Device Identifier: ABL90 (see attached)
Reason of Field Safety Corrective Action: The linearity of the glucose is dependent on the oxygen tension of the sample. This dependence is due to the co-reaction of glucose and oxygen by the enzyme glucose oxidase. Very low pO2 levels can influence the linearity of the glucose sensor. As a consequence of this Glu results will no longer be reported when the pO2 value is below 25 mmHg.
Remedy Action: Upgrade the software in all ABL90 FLEX analyzers to V2.8. Provide each customer with the customer letter as well as a copy of the Update kits for the Operator’s and Reference Manuals.
Athorized Representative/Importer/Distributor:Salehiya Trading Est., Riyadh, (01) 2628939
Report Source:NCMDR
Source Ref. Number:9BB13938BA277
SFDA Comments:SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Attachments: Radiometer Medical (3).pdf