NCMDR Recall
Back
Reference Number:
mdprc 001 08 12 000
Date submitted:
8/1/2012
Manufacturer:
Abbott Laboratories Inc
Device Type:
IVD, ARCHITECT Toxo IgG reagent
Description:
IVD, reagent
Medical Device Identifier:
List Number: 6C19-25 Lot Numbers: 15070LI00 and 15070LI01
Reason of Field Safety Corrective Action:
ARCHITECT Toxo IgG reagent are showing reduced specificity which may cause a shift to higher IU/mL values leading to increased grayzone and/or reactive results.
Remedy Action:
Determine if you are currently using or have inventory of lots 15070LI00 and/or 15070LI01, immediately contact Customer Support for replacement material. Discontinue use of lots 15070LI00 and/or 15070LI01 and destroy any remaining inventory of these lots according to your laboratory procedures.
Athorized Representative/Importer/Distributor:
Medical supplies & Services Co.Ltd Mediserv, Riyadh, (01) 4780555
Report Source:
NCMDR
Source Ref. Number:
1FCA1F5CD0234
SFDA Comments:
SFDA urges all hospitals that have devices subjected to recall, to contact the company.