NCMDR Recall
Back
Reference Number:
mdprc 035 07 12 000
Date submitted:
7/15/2012
Manufacturer:
Stryker Instruments
Device Type:
Instrument Trays, Arthroscope and Hardware Set Trays
Description:
Instrument Trays, Sterilization Containers
Medical Device Identifier:
Part No. 233-032-116 All lot
Reason of Field Safety Corrective Action:
During a review and retest of the sterilization parameters, gravity steam, EO and Sterrad sterilization cycles did not consistently sterilize all of the products in the tray.
Remedy Action:
Identify any affected product in your inventory. Ensure that any obsolete IFUs are discard, and find attached for new IFU.
Athorized Representative/Importer/Distributor:
Al-Faisaliah Medical System, Riyadh, (01) 2119948
Report Source:
NCMDR
Source Ref. Number:
B27F6F86BB2E1
SFDA Comments:
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Attachments:
Autoclavable Arthroscope Hardware Tray.pdf
View History