Reference Number: mdprc 104 03 12 000
Date submitted: 3/25/2012
Manufacturer:GE Medical Systems
Device Type:Optima MR450W with GEM option, MRI
Description:MRI, systems, RF
Medical Device Identifier:Model number: 5410917 MR450W
Reason of Field Safety Corrective Action:During scanning of a patient, if an improperly torqued gradient connection exists, it may cause the gradient lead to produce excessive heat. And if a second and compounding issue of improperly routed Rear Pedestal Air Supply Hose is also present, routed in such a way to cause the hose to come in contact with the overheated gradient lead cable, then the hose may generate smoke.
Remedy Action:Field action to inspect (and correct if needed) the routing of the air hose of the system to ensure that it is routed per the existing service documentation.
Athorized Representative/Importer/Distributor:GE Healthcare, Riyadh, (01) 4600530
Report Source:NCMDR
Source Ref. Number:954D7D34631F6
SFDA Comments:SFDA urges all hospitals that have devices subjected to recall, to contact the company.