Reference Number:
mdprc 036 01 19 000
Date submitted:
1/14/2019
Manufacturer:
Varian Medical Systems
Device Type:
ARIA Oncology Information System for Radiation Oncology
Description:
Radiation Oncology
Medical Device Identifier:
Affected Version(s) / Lot(s): 15.1, 15.5, 15.6
Reason of Field Safety Corrective Action:
Varian Medical Systems has identified a problem in ARIA Oncology Information System (OIS) for Radiation Oncology such that a user may inadvertently chart medical record information in an unintended patient’s chart.
Remedy Action:
-Varian recommends that Users close each patient prior to opening a new one. This will prevent this issue from occurring.
-Varian Medical Systems is investigating this issue. The timeline of a technical solution is currently unknown. Varian will contact users as a fix is available.
Athorized Representative/Importer/Distributor:
Varian Medical Systems Arabia Commercial Limited
Report Source:
NCMDR
Source Ref. Number:
138B3678601AC
SFDA Comments:
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Attachments:
Varian Medical Systems.pdf