X
OK Your changes have been successfully saved
SFDA Home Page
Medical Devices Sector
قطاع الأجهزة الطبية
NCMDR
National Center for Medical Devices Reporting
المركز الوطني لبلاغات الأجهزة والمنتجات الطبية
NCMDR Recall
Reference Number: mdprc 086 11 17 001
Date submitted: 3/7/2018
Manufacturer:Cordis Corporation
Device Type:Cordis S.M.A.R.T.® Flex Vascular Stent System.
Description:Flex Vascular Stent System
Medical Device Identifier:85 lot numbers of 48 catalog numbers. Reference Table 1. The lot number range is from 40932 through 41637, but not all lots in the range are affected.


..................................
NCMDR Update Code:
GTIN: Catalog No.
20705032066607: SF06100SV
20705032066805: SF07200SV
20705032066591: SF08060SV
20705032066768: SF08100MV
20705032067192: SF08100SV
20705032066645: SF10040SV
Multiple Lot numbers of the affected device are provided in the attached FSN
Reason of Field Safety Corrective Action:Based on one complaint and the subsequent investigation, Cordis has determined that 85 distributed lots of S.M.A.R.T.® Flex Vascular Stent System have a potential for cracked luer hubs, due to a manufacturing error.
The luer hub is used to inject contrast/saline during procedure preparation, prior to insertion of the stent system into the patient. A cracked hub could result in leakage of saline, or in the worst case, inability to flush the outer sheath of the catheter.
Remedy Action:Check your inventory to determine if you have any remaining affected product in your possession. Check all storage and usage locations. The purpose is to identify the product, not to remove the product.
During procedure preparation, check for cracks/leaks. If the user detects a crack prior to flushing, or a leak during flushing, select a replacement device. (See inspection section below.) Report the incident through the standard complaint process.
Athorized Representative/Importer/Distributor:ABDULREHMAN AL GOSAIBI GTB
Report Source:NCMDR
Source Ref. Number:83692649551B0
SFDA Comments:SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Attachments: Cordis Corporation.pdf