Reference Number:
mdprc 008 02 11 000
Date submitted:
2/2/2011
Manufacturer:
Nihon Kohden Corp
Device Type:
KC 600P Cart, beside Monitor
Description:
Monitors, Bedside, Physiologic
Medical Device Identifier:
WEP 5200 Telemetry System, BSM 6000 series or PVM 2701 Beside Monitor attached to KC 600P Cart
Reason of Field Safety Corrective Action:
A fault in the lock plate of the cart, some lock plate are not thick enough to support the monitor under condition of frequent tilting. In a worst-case situation, the monitor can fall over
Remedy Action:
Replace the lock plate with the provided one. Follow the procedure in the installation guide included with the provided lock plate
Athorized Representative/Importer/Distributor:
ABDULREHMAN AL GOSAIBI GTB, Riyadh (01) 4793000
Report Source:
NCMDR
Source Ref. Number:
2773FC4828206
SFDA Comments:
SFDA urges all hospitals that have devices subjected to recall, to contact the company