X
Your changes have been successfully saved
Medical Devices Sector
قطاع الأجهزة الطبية
NCMDR
N
ational
C
enter for
M
edical
D
evices
R
eporting
المركز الوطني لبلاغات الأجهزة والمنتجات الطبية
Home
Published FSNs/Recalls
About NCMDR
Contact Us
FAQ
Login
NCMDR Recall
Back
Reference Number:
mdprc 126 10 16 000
Date submitted:
10/19/2016
Manufacturer:
Medtronic SA
Device Type:
Meddtronic Model 37751 Recharger included in Model 37651 Charging System
Description:
Neuromuscular electrical stimulation system programmer, Battery charger
Medical Device Identifier:
Model 37651, 37754 and 97754 (containing the model 37751 recharger)
Reason of Field Safety Corrective Action:
Medtronic has identified an increased number of complaints from customers involving reports of Rechargers that are in an unresponsive error state, where the Recharger is non-functional with a blank display screen and is beeping every 5 seconds.
Remedy Action:
Please check out the recommendations in the attached FSN.
Athorized Representative/Importer/Distributor:
Medtronic Saudi Arabia
Report Source:
NCMDR
Source Ref. Number:
7E665AF6A52D9
SFDA Comments:
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Attachments:
Medtronic Inc..pdf