X
OK Your changes have been successfully saved
SFDA Home Page
Medical Devices Sector
قطاع الأجهزة الطبية
NCMDR
National Center for Medical Devices Reporting
المركز الوطني لبلاغات الأجهزة والمنتجات الطبية
NCMDR Recall
Reference Number: mdprc 028 07 11 000
Date submitted: 7/9/2011
Manufacturer:Varian Medical Systems
Device Type:Treatment planning system, Eclipse
Description:Radiation Therapy
Medical Device Identifier:Eclipse Release 6.5 through 10 (build 7.3.10 through 10.0.34)
Reason of Field Safety Corrective Action:In affected Eclipse versions, user editing of MLC leaf positions may not automatically invalidate calculated dose distributions.
Remedy Action:Varian is preparing a maintenance release (service pack) of Eclipse software to address this issue. You will be contacted by a Varian representative to schedule installation of a corrected version once available.
Athorized Representative/Importer/Distributor:Varian Medical Systems, Riyadh, (01) 2772126
Report Source:NCMDR
Source Ref. Number:AD9DDB80042A9
SFDA Comments:SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Attachments: Varian Medical Systems - FSN.pdf