Reference Number:
mdprc 098 01 16 000
Date submitted:
1/28/2016
Manufacturer:
Leica Microsystems Inc.
Device Type:
Leica M525 OH4 and Leica M720 OH5
Description:
Surgical microscope systems
Medical Device Identifier:
Leica OH4/ OH5 Surgical Microscope Stands with 400W illumination manufactured from 1st October 2012 until 1st September 2014
Multiple serial numbers affected
Reason of Field Safety Corrective Action:
An electronic component in one of the two illumination units can potentially overheat, resulting in minimal smoke and odor emanating from the ventilation openings of the illumination units.
Remedy Action:
A Leica Microsystems representative will be contacting your facility to arrange the no-cost replacement of the power supply output cables with improved cables including a filter coil in both 400W illumination units of your instrument as soon as possible.
Athorized Representative/Importer/Distributor:
Al-Faisaliah Medical System
Report Source:
NCMDR
Source Ref. Number:
34786AAF50215
SFDA Comments:
SFDA urges all hospitals that have devices subjected to recall, to contact the company.