NCMDR Recall
Back
Reference Number:
mdprc 004 11 15 001
Date submitted:
11/8/2015
Manufacturer:
B Braun Medical Inc
Device Type:
A/V Set DEHP-free PVC for Dialog
Description:
Dialysis, A/V Set
Medical Device Identifier:
Art. No. 7211034 Batch # 15C20980
……………………………
BfArM update code :
7211032 : 15C16980
7211033 : 15C23980
7211035 : 15C16980
7210868 : 15D27880
Reason of Field Safety Corrective Action:
During continuous post-market surveillance we became aware of a disconnection of the female Luer-Lock connector from the venous side of the Single-Needle adapter leading to blood loss for the patient.
Remedy Action:
-Check whether you have above mentioned products in stock, and quarantine them. The isolated products will be exchanged according to your information on the fax form. -At the next delivery the quarantined products will be exchanged according to your information given on the return fax
Athorized Representative/Importer/Distributor:
Medical supplies & Services Co.Ltd Mediserv
Report Source:
BfArM
Source Ref. Number:
7419/15
SFDA Comments:
SFDA urges all hospitals that have devices subjected to recall, to contact the company.