Back
Reference Number:
mdprc 083 10 14 000
Date submitted:
10/21/2014
Manufacturer:
Mortara Instrument, Inc.
Device Type:
Mortara Surveyor S12 and Surveyor S19 Patient Monitors
Description:
Patient Monitor
Medical Device Identifier:
Product Name : SURVEYOR Mortara Part Number : SUR12-BCF-AXXAD Device Serial Number : 114170207572
Product Name : SURVEYOR Mortara Part Number : SUR19-BCF-AXXAC Device Serial Number : 114170207577
Reason of Field Safety Corrective Action:
A hazardous situation may be created when mounting these monitors using the 75mm VESA mounting hole pattern provided in the battery cover of these monitors.
Remedy Action:
A correction has been developed to eliminate the potential risk. The correction replaces the original battery cover with a new design incorporating a barrier that prevents mounting screws from penetrating into the battery compartment.
Athorized Representative/Importer/Distributor:
Saudi Health Services Co. Ltd.
Report Source:
NCMDR
Source Ref. Number:
8EB7F7CCFB403
SFDA Comments:
SFDA urges all hospitals that have devices subjected to recall, to contact the company.