X
OK Your changes have been successfully saved
SFDA Home Page
Medical Devices Sector
قطاع الأجهزة الطبية
NCMDR
National Center for Medical Devices Reporting
المركز الوطني لبلاغات الأجهزة والمنتجات الطبية
NCMDR Recall
Reference Number: mdprc 083 10 14 000
Date submitted: 10/21/2014
Manufacturer:Mortara Instrument, Inc.
Device Type:Mortara Surveyor S12 and Surveyor S19 Patient Monitors
Description:Patient Monitor
Medical Device Identifier:Product Name : SURVEYOR Mortara Part Number : SUR12-BCF-AXXAD Device Serial Number : 114170207572

Product Name : SURVEYOR Mortara Part Number : SUR19-BCF-AXXAC Device Serial Number : 114170207577
Reason of Field Safety Corrective Action:A hazardous situation may be created when mounting these monitors using the 75mm VESA mounting hole pattern provided in the battery cover of these monitors.
Remedy Action:A correction has been developed to eliminate the potential risk. The correction replaces the original battery cover with a new design incorporating a barrier that prevents mounting screws from penetrating into the battery compartment.
Athorized Representative/Importer/Distributor:Saudi Health Services Co. Ltd.
Report Source:NCMDR
Source Ref. Number:8EB7F7CCFB403
SFDA Comments:SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Attachments: Mortara .pdf