Reference Number: mdprc 032 07 13 000
Date submitted: 7/3/2013
Manufacturer:PAUL HARTMANN
Device Type:Folio dress gown Comfort Basic
Description:sterile, Single-use
Medical Device Identifier:See attached
Reason of Field Safety Corrective Action:A quality inspection revealed the individual occurrence of micro-holes in a number of blisters (sterile packaging) containing terile surgical gowns and isolation gowns. The micro-holes are unlikely to be detected through visual inspection. As a result, the sterility of affected products may be compromised.
Remedy Action:Please examine your inventory and do not use the affected products any longer and return product
Athorized Representative/Importer/Distributor:ABDULREHMAN AL GOSAIBI GTB, Riyadh, (011) 4771374
Report Source:NCMDR
Source Ref. Number:FCCA25EF2D307
SFDA Comments:SFDA urges all hospitals that have devices subjected to recall, to contact the company
Attachments: PAUL.pdf