NCMDR Recall
Reference Number: mdprc 136 09 12 000
Date submitted: 9/16/2012
Manufacturer:Philips Healthcare
Device Type:HeartStart XL Defibrillator/Monitor Battery
Description:Defibrillator/Monitor
Medical Device Identifier: Units Affected: HeartStart XL battery M3516A labeled with Made in Taiwan and an R-2011-12 Date of Manufacture.
Reason of Field Safety Corrective Action: If the XL Defibrillator/Monitor (M4735A) is used on a patient when AC power is not in use, with a battery that may fail, the device may shut down unexpectedly without providing the user with a low battery warning message or audible tone.
Remedy Action: Philips will Providing affected customers no charge replacement battery (s) as soon as new production is available. A Philips Healthcare representative will contact customers with affected battery(s) to arrange for replacement.
Athorized Representative/Importer/Distributor:Al-Faisaliah Medical System, Riyadh, (01) 2119948
Report Source:NCMDR
Source Ref. Number:2329CD26981D7
SFDA Comments:SFDA urges all hospitals that have devices subjected to recall, to contact the company.