Reference Number:
mdprc 029 01 19 000
Date submitted:
1/10/2019
Manufacturer:
Philips Healthcare
Device Type:
Philips PageWriter TC Cardiographs (TC20/30/50/70)
Description:
Electrocardiographs, Multichannel
Medical Device Identifier:
Product Codes: 860315, 860310, 860306, 860332, 860352, 860353, 860354, 860355, 860429
Reason of Field Safety Corrective Action:
Philips has received a report in which the lithium ion battery in a Philips PageWriter TC cardiograph overheated and ignited.
Remedy Action:
Philips plans to release a customer installable, software update for PageWriter TC cardiograph (TC20/30/50/70) that will provide alerts to assist users in managing the battery replacement cycle. You will be notified when the customer installable software update is available.
Athorized Representative/Importer/Distributor:
Philips Healthcare Saudi Arabia Ltd.
Report Source:
NCMDR
Source Ref. Number:
5BACF7BD122CD
SFDA Comments:
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Attachments:
Philips Healthcare.pdf
View History