NCMDR Recall
Back
Reference Number:
mdprc 062 04 18 000
Date submitted:
4/11/2018
Manufacturer:
Siemens Healthcare Diagnostics GmbH
Device Type:
ADVIA Chemistry Systems Urea Nitrogen (UN)
Description:
Urea Nitrogen
Medical Device Identifier:
Catalog number: 03040257
Siemens Material Number: 10309051
Lot Number: 408143
Exp Date 28/06/2018
Lot Number: 418290
Exp Date 28/06/2018
Reason of Field Safety Corrective Action:
Siemens Healthcare Diagnostics has confirmed that the Urea Nitrogen reagent kit lots listed in the table 1 above for use on ADVIA® 1800, 2400 and XPT Chemistry Systems demonstrate an increased incidence of Calibration Failures. Results cannot be generated by the system when calibration fails.
If calibration passes, there is a potential for falsely elevated patient and quality control results.
Remedy Action:
Discontinue use of and discard the kit Lots listed above.
Athorized Representative/Importer/Distributor:
ABDULREHMAN AL GOSAIBI GTB
Report Source:
NCMDR
Source Ref. Number:
DDDCC8C9EB435
SFDA Comments:
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Attachments:
Siemens Healthcare Diagnostics.pdf