X
OK Your changes have been successfully saved
SFDA Home Page
Medical Devices Sector
قطاع الأجهزة الطبية
NCMDR
National Center for Medical Devices Reporting
المركز الوطني لبلاغات الأجهزة والمنتجات الطبية
NCMDR Recall
Reference Number: mdprc 160 08 15 000
Date submitted: 8/24/2015
Manufacturer:Medtronic Inc
Device Type:TRITON™ Pneumatic High Torque Hand Piece Hose
Description:High Torque Hand Piece Hose
Medical Device Identifier:Model Number: 720101 Serial Numbers: 152369, 152377, 152379, 152380
Reason of Field Safety Corrective Action:The reason for the removal is that MicroAire Surgical Instruments has detected potential workmanship issues that may result in a decreased flow of nitrogen through the hose.
Remedy Action:Return all potentially affected product in your inventory to Medtronic. Your Medtronic Representative will assist you with this process.
Athorized Representative/Importer/Distributor:Gulf Medical Co.
Report Source:NCMDR
Source Ref. Number:A18AE5B2202E2
SFDA Comments:SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Attachments: Medtronic Inc.pdf