NCMDR Recall
Back
Reference Number:
mdprc 091 11 11 000
Date submitted:
11/27/2011
Manufacturer:
Varian Medical Systems
Device Type:
Radiation Therapy, Eclipse with ARIA™ Radiation Oncology (RT Chart module)
Description:
Radiation Therapy, treatment planning system
Medical Device Identifier:
Eclipse Release 10 with ARIA™ Radiation Oncology (RT Chart module) Release 10 (Client Build 10.0.28 or Build 10.0.34). Serial numbers list attached.
Reason of Field Safety Corrective Action:
An anomaly has been identified within Varian Eclipse™ when used with ARIA™ Radiation Oncology (RT Chart module) where adding a reference point in RT Chart and editing the field dose contributions may change the Prescribed Dose Per Fraction in Eclipse.
Remedy Action:
Varian will be notifying all affected customers with a letter. This issue has been corrected in a maintenance release (10MR2, Client Build 10.0.39) of Eclipse and ARIA RT Chart. All affected customers will be contacted by Varian to schedule installation.
Athorized Representative/Importer/Distributor:
Varian Medical Systems, Riyadh, (01) 2772126
Report Source:
NCMDR
Source Ref. Number:
2B8A6FE5982A1
SFDA Comments:
SFDA urges all hospitals that have devices subjected to recall, to contact the company.