NCMDR Recall
Back
Reference Number:
mdprc 187 02 12 000
Date submitted:
2/29/2012
Manufacturer:
GE Healthcare
Device Type:
X-ray system, diagnostic. Series 7600/7700 Mobile C-Arm,, Interconnect PO Cable
Description:
X-ray system, diagnostic, fluoroscopic, general purpose, mobile, digital
Medical Device Identifier:
Series 7600/7700 Mobile C-Arm, Compact 7600/7700 Mobile C-Arm, Compact 7700 Plus Mobile C-Arm. Used with interconnect PO Cable part number: 00-452118-01 or 00-453118-01. Serial Number range: 2457-2632.
Reason of Field Safety Corrective Action:
The interconnect PO Cable that connects the C-arm to the monitor cart may fail, resulting in loss of video signal transmission to the monitors.
Remedy Action:
Avoid making sharp or low angle bends with the PO interconnect cable. GE Healthcare will arrange for replacement of the interconnect cable at no charge.
Athorized Representative/Importer/Distributor:
GE Healthcare, Riyadh, (01) 4600530
Report Source:
NCMDR
Source Ref. Number:
AA38391DA91E1
SFDA Comments:
SFDA urges all hospitals that have devices subjected to recall, to contact the company.