NCMDR Recall
Reference Number: mdprc 187 02 12 000
Date submitted: 2/29/2012
Manufacturer:GE Healthcare
Device Type:X-ray system, diagnostic. Series 7600/7700 Mobile C-Arm,, Interconnect PO Cable
Description:X-ray system, diagnostic, fluoroscopic, general purpose, mobile, digital
Medical Device Identifier:Series 7600/7700 Mobile C-Arm, Compact 7600/7700 Mobile C-Arm, Compact 7700 Plus Mobile C-Arm. Used with interconnect PO Cable part number: 00-452118-01 or 00-453118-01. Serial Number range: 2457-2632.
Reason of Field Safety Corrective Action:The interconnect PO Cable that connects the C-arm to the monitor cart may fail, resulting in loss of video signal transmission to the monitors.
Remedy Action:Avoid making sharp or low angle bends with the PO interconnect cable. GE Healthcare will arrange for replacement of the interconnect cable at no charge.
Athorized Representative/Importer/Distributor:GE Healthcare, Riyadh, (01) 4600530
Report Source:NCMDR
Source Ref. Number:AA38391DA91E1
SFDA Comments:SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Attachments: GE Healthcare.pdf