NCAR Recall
Reference Number: mdprc 010 06 12 000
Date submitted: 6/2/2012
Manufacturer:ThermoFisher Scientific Microbiology Perth
Device Type:Oxoid Antimicrobial Susceptibility Testing Disc
Description:IVD, Oxoid, Testing Disc
Medical Device Identifier:ATM30 (Aztreonam), CT0264B, Lot 1100699
Reason of Field Safety Corrective Action:Cartridges of this lot of Oxoid Antimicrobial Susceptibility Testing may contain individual discs that are not impregnated with antibiotic.
Remedy Action:Destroy any remaining inventory of the lot listed above and contact Customer Services or your local Oxoid supplier who will issue a replacement
Athorized Representative/Importer/Distributor:Fuad Abdul Jalil Al Fadhli & Sons, Riyadh, (01) 4550807
Report Source:NCAR
Source Ref. Number:UK-2012-05-22-099
SFDA Comments:SFDA urges all hospitals that have devices subjected to recall, to contact the company.